TB 9-6625-2277-35
Test
Calibrator
err
instrument
Calibrator
display indications
range
output
(%)
pushbuttons
nominal value
200 Ω
100 Ω
.3
1 kΩ
2
.3
10 kΩ
20
.3
200
100 kΩ
.3
1 MΩ
2000
.6 (.3)1
20 MΩ
10 MΩ
.6
1Model
8010M.
NOTE
Perform (4) through (10) below for models 8010A and 8010A-01.
(4) Simultaneously press 200Ω and 2 range pushbuttons to select 2 mS range.
(5) Set calibrator for 1000 Ω nominal output with 2 wire comp ON. TI will
indicate between .997 and 1.003.
(6) Simultaneously press 20 and 200 range pushbuttons to select 20 S range.
(7) Set calibrator for 100 kΩ nominal output. TI will indicate between 9.97 and 10.03.
(8) Simultaneously press 2000 and 20MΩ range pushbuttons to select 200 nS range.
(9) Set calibrator for 10 MΩ nominal output. TI will indicate between 98.0 and 102.0.
b. Adjustments. No adjustments can be made.
12. Resistance (Model 8012A only)
a. Performance Check
(1) Connect calibrator OUTPUT HI to TI LO RANGE Ω and calibrator OUTPUT
LO to TI COMMON.
(2) Simultaneously press V and mA function pushbuttons to select LO RANGE Ω
function and press 2 range pushbutton.
NOTE
Calibrator 2 wire Comp must be set to ON in (3) through (5)
below.
(3) Set calibrator output for 0 ohm. Adjust TI ZERO control for a .000 to .001 TI
indication.
(4) Set calibrator for a nominal 1.9 ohm output. Rotate calibrator knob below EDIT
FIELD pushbutton to adjust calibrator display indication to equal TI indication. If
calibrator err display does not indicate within 1.105%, perform b below.
(5) Press 20 range pushbutton.