TB 9-6625-2280-35
(3) Adjust signal generator amplitude controls for a power meter indication of 10.0 mW.
(4) Disconnect power meter from POINT A (fig. 2) and 50 Ω termination from
(5) Connect power meter to POINT B (fig. 2) and 50 Ω termination to POINT A (fig. 2).
(6) Record power meter indication as P1 value.
(7) Disconnect power meter from POINT B (fig. 2) and 50 Ω termination from
(9) Adjust signal generator amplitude controls for a power meter indication of 1.0 mW.
(10) Disconnect power meter from POINT A (fig. 2) and 50 Ω termination from
(11) Connect power meter to POINT B (fig. 2) and 50 Ω termination to POINT A (fig. 2).
(12) Record power meter indication as P2 value.
(13) Disconnect 50 Ω termination from POINT A (fig. 2).
(14) Divide P2 value recorded in (12) above into P1 value recorded in (6) above.
Record results as X.
(15) Press TI keys as listed in (a) through (d) below:
(a)
MENU key (two times).
(b)
F1 key.
(c)
500 using number keys.
Any UNITS key (nS or S or mS key).
(d)
NOTE
If TI display does not indicate mW, perform (16) below.
(16) Press MENU key four times and then press F3 key to place TI EL display
indication in mW.
NOTE
This action changes menu format from dBm to mW.
(17) Connect RF detector A to CALIBRATOR 1 GHZ output.
(18) Press keys as listed in (a) through (d) below:
(a)
MENU key.
(b)
F3 key.
(c)
F1 key.
Any UNITS key (nS or S or mS key).
(d)
(19) Disconnect RF detector A from CALIBRATOR 1 GHZ output.
NOTE
10