TB 9-6625-2280-35
(c) F2 key.
(12) Adjust A6R112 (A7R112) (fig. 3) for an oscilloscope display indication between
-0.002 and +0.002 V dc offset at the end of sample time (R).
NOTE
Press channel B key in (13) (c) below when repeating steps for
A7 board adjustments.
(13) Press TI keys as listed in (a) through (e) below:
(a)
MENU key.
(b)
F1 key.
(c)
Channel A key (channel B key).
(d)
-15 using number keys.
(e)
dBm key.
(14) Connect multimeter (dc mode) LO lead to A6TP1 (A7TP1) (fig. 3).
(15) Connect multimeter (dc mode) HI lead to A6TP13 (A7TP13) (fig. 3).
(16) Adjust A6R113 (A7R113) (fig. 3) for a multimeter indication between -0.001 and
+0.001 V dc offset (R).
(17) Disconnect TI from equipment setup.
(18) Repeat (6) through (17) above for A7 board.
(19) Press RESET pushbutton located on TI rear panel.
(20) Press TI keys as listed in (a) through (d) below:
(a)
MENU key (eleven times).
(b)
F1 key.
(c)
1 using number key.
(d)
F2 key.
NOTE
"CONNECT DETECTOR TO CALIBRATOR THEN PRESS
ANY UNITS KEY", perform the instructions and then press 7
data entry key.
(21) Connect multimeter (dc mode) to ANALOG A output on TI rear panel.
(22) Adjust A6R47 (fig. 3) for a multimeter indication between -0.01 and +0.01 V dc (R).
(23) Press F3 key. Multimeter will indicate between 2.975 and 3.025 V dc.
(24) Press F1 key. Multimeter will indicate between -4.965 and -5.035 V dc.
(25) Press F2 key.
(26) Move multimeter (dc mode) connection from ANALOG A output on TI rear
panel to ANALOG B output on TI rear panel.
(27) Adjust A7R47 (fig. 3) for a multimeter indication between -0.01 and +0.01 V dc (R).
16