TB 9-6625-2282-35
5. Accessories Required. The
accessories required for the
calibration are common usage
accessories, issued as indicated in paragraph 4 above and are not listed in this calibration
procedure.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
(13335470)
NIGHT VISION DEVICE
Wavelength range: 305 to 1000 nm
DETECTOR STANDARD
Wavelength peak response: 720 nm +50
Responsitivity: 1.51 x 107
V/W sr-1 at 810 nm +8% (+11%)
Dynamic range: 6.6 x 10-13 to 4.6 x 10-7
Gain (elect): 109 V/A
Hewlett Packard, Model 3458A
Range: 0.62 to 732 mV
(3458A)
Resolution: .01 mV
Accuracy: +1.6%
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in TM 11-5855-299-12&P.
d. Unless otherwise specified, all controls and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Remove TI from protective case.
b. Ensure TI II, III, OFF switch is set to OFF position.
c. Set multimeter to
measure dc
volts and
auto-ranging.