TB 9-6625-2288-24
(19) Adjust calibrator for 0.003 V ac output. Adjust 10MV LO (fig. 1) for a meter
indication between 0.0025 and 0.0035 V ac (ideal 0.003) (R).
(20) Repeat (17) through (19) above until no further adjustments are required.
(21) Set TI RANGE-FULL-SCALE switch to .03-VOLTS.
(22) Adjust calibrator for 0.03 V ac output. Adjust 30MV HI (fig. 1) for a meter
indication between 0.0285 and 0.0315 V ac (ideal 0.03) (R).
(23) Adjust calibrator for 0.02 V ac output. Adjust 30MV MID (fig. 1) for a meter
indication between 0.0185 and 0.0215 V ac (ideal 0.02) (R).
(24) Adjust calibrator for 0.01 V ac output. Adjust 30MV LO (fig. 1) for a meter
indication between 0.0085 and 0.0115 V ac (ideal 0.01) (R).
(25) Repeat (22) through (24) above until no further adjustments are required.
(26) Set TI RANGE-FULL-SCALE switch to .1-VOLTS.
(27) Adjust calibrator for 0.1V ac output. Adjust .1V HI (fig. 1) for a meter indication
between 0.095 and 0.105 V ac (ideal 0.1) (R).
(28) Adjust calibrator for 0.06 V ac output. Adjust .1V MID (fig. 1) for a meter
indication between 0.055 and 0.065 V ac (ideal 0.06) (R).
(29) Adjust calibrator for 0.03 V ac output. Adjust .1V LO (fig. 1) for a meter
indication between 0.025 and 0.035 V ac (ideal 0.03) (R).
(30) Repeat (27) through (29) above until no further adjustments are required.
(31) Set TI RANGE-FULL-SCALE switch to .3-VOLTS.
(32) Adjust calibrator for 0.3 V ac output. Adjust .3V HI (fig. 1) for a meter
indication between 0.285 and 0.315 V ac (ideal 0.3 V) (R).
(33) Adjust calibrator for 0.2 V ac output. Adjust .3V MID (fig. 1) for a meter
indication between 0.185 and 0.215 V ac (ideal 0.2) (R).
(34) Adjust calibrator for 0.1V ac output. Adjust. .3V LO (fig. 1) for a meter
indication between 0.085 and 0.115 V ac (ideal 0.1) (R).
(35) Repeat (32) through (34) above until no further adjustments are required.
(36) Set TI RANGE-FULL-SCALE switch to 1 -VOLTS.
(37) Adjust calibrator for 1 V ac output. Adjust 1V HI (fig. 1) for a meter indication
between 0.950 and 1.05 V ac (ideal 1) (R).
(38) Adjust calibrator for 0.6 V ac output. Adjust 1V MID (fig. 1) for a meter
indication between 0.550 and 0.650 V ac (ideal 0.6) (R).
(39) Adjust calibrator for 0.3 V ac output. Adjust 1V LO (fig. 1) for a meter
indication between 0.250 and 0.350 V ac (ideal 0.3) (R).
(40) Repeat (37) through (39) above until no further adjustments are required.
(41) Set TI RANGE-FULL-SCALE switch to 3-VOLTS.
(42) Adjust calibrator for 3 V ac output. Adjust 3V HI (fig. 1) for a meter indication
between 2.85 and 3.15 V ac (ideal 3) (R).
7