Accuracy and Linearity
Power Supply

Calibration Procedure For Electronic Voltmeters An/Urm-145D (Me-247D/U) (Model Mv-828A) Page Navigation
  2    3    4    5    6  7  8    9  
img
. . . . . . . . . .
TB 9-6625-2288-24
(43) Adjust calibrator for 2 V ac output. Adjust 3V MID (fig. 1) for a meter indication
between 1.85 and 2.15 V ac (ideal 2) (R).
(44) Adjust calibrator for 1 V ac output. Adjust 3V LO (fig. 1) for a meter indication
between 0.850 and 1.15 V ac (ideal 1) (R).
(45) Repeat (42) through (44) above until no further adjustments are required.
(46) Disconnect TI RF probe from calibrator.
9. Frequency Response
a. Performance Check
(1) C o n n e c t c a l i b r a t o r W I D E B A N D O U T P U T ( w i t h o u t 5 0 o h m p a d ) t o T I
R F probe using 50 ohm feed through termination supplied with TI.
(2) Set RANGE-FULL-SCALE switch to .1-VOLTS.
(3) Adjust calibrator frequency controls to value listed in table 4.
(4) Adjust calibrator output voltage controls for 0.1 V indication on TI meter.
Calibrator voltage output will be within the limits specified in table 4.
(5) Repeat (3) and (4) above for remaining calibrator frequency settings listed in
table 4.
Table 4. Frequency Response
Calibrator
Calibrator output voltage
frequency settings
(mV)
(kHz)
Min
(Max)
20
95
105
50
95
105
100
95
105
500
95
105
1000
95
105
(6) Connect equipment as shown in figure 2.
Figure 2. Frequency response - equipment setup.
8