TB 9-6625-2289-35
(15) Adjust METER CAL R49 (fig. 1) for a TI indication between 2.7 and 3.3 V ac
(ideal 3.0 V ac) (R).
(16) Set RANGE switch to .001 VOLTS.
(17) Disconnect RF probe from calibrator.
(18) Zero TI meter with ZERO control.
NOTE
If necessary, repeat 7 g through i above.
(19) Connect probe to calibrator WIDEBAND OUTPUT.
(20) Adjust calibrator for 1 mV ac output and then adjust R3 (fig. 1) for a TI
indication between .00085 and .001 15 V ac (ideal .001 V ac) (R).
(21) Set RANGE switch to .003 VOLTS.
(22) Disconnect RF probe from calibrator.
(23) Zero TI meter with ZERO control.
NOTE
If necessary, repeat 7 g through i above.
(24) Connect RF probe to calibrator WIDEBAND OUTPUT.
(25) Adjust calibrator for 3 mV ac output and then adjust R6 (fig. 1) for a TI
indication between .0027 and .0033 V ac (ideal .003 V ac) (R).