SECTION III. CALIBRATION PROCESS
Transmit Verification

Calibration Procedure For Optical Attenuation Test Set Ts-4335/G Page Navigation
  1    2  3  4    5    6  
img
. . .
TB 9-6625-2291-24
NOTE
Substitution of lower grade isopropyl alcohol or freon based
cleaners can cause damage to connectors.
b. Ni-Cad rechargeable batteries should be in fully charged state. Push PULL ON
pushbutton to (off), charge 14 hours minimum.
c. Pull PULL ON pushbutton to (on). Ensure TI display has clear crisp resolution, if
not adjust A2R12 (fig. 1) for best clarity and resolution.
NOTE
All adjustments are made on measurement and display printed
circuit board (fig. 1).
A2R9
U3
A2R43
A2R12
A2R3
A2R48
Figure 1. Measurement and display printed circuit board.
4