TB 9-6625-2293-35
Test
Test instrument trigger level readout indications
instrument
- Peak
VOLTS/DIV
+ Peak
switch
output
settings
settings
Min
Max
Min
Max
2
mV
10
mV
8.5 mV
11.5 mV
-1.2 mV
1.2 mV
5
mV
20
mV
17.3 mV
22.7 mV
-2.1 mV
2.1 mV
10
mV
50
mV
44.5 mV
55.5 mV
-4.0 mV
4.0 mV
20
mV
0.1 V
89
mV
111
mV
-7.5 mV
7.5 mV
50
mV
0.2 V
177
mV
223
mV
-17
mV
17
mV
100
mV
0.5 V
449
mV
551
mV
-36
mV
36
mV
200
mV
1.0 V
0.90 V
1.10 V
-70
mV
70
mV
500
mV
2.0 V
1.78 V
2.22 V
-.14 V
.14 V
1.0 V
5.0 V
4.50 V
5.50 V
-.35 V
.35 V
2.0 V
10.0 V
9.0 V
11.0 V
-0.7 V
0.7 V
5.0 V
20.0 V
17.8 V
22.2 V
-1.4 V
1.4 V
(8) Set oscilloscope calibrator for a CHAN 1, VOLTAGE mode output of 10 mV at 1
kHz frequency and set TI CH 1 VOLTS/DIV switch to 2 mV.
(9) Pull SEC/DIV knob out and press A/B TRIG pushbutton for B TRIGGER and
push TRIGGER MODE down for TRIG AFT DLY. Adjust TRIGGER LEVEL for a
stable display.
(10) Adjust ∆REF OR DLY POS control for a delay readout of 0.000 ms.
(11) Adjust TRIGGER LEVEL control for most positive voltage that produces an
intensified point on waveform display for both + and - SLOPE. If B TRIGGER level
readout indication is not within limits given in + peak column of table 7 for 2 mV setting,
perform b below.
(12) Adjust TRIGGER LEVEL control for most negative voltage that produces an
intensified point on waveform display for both + and - SLOPE. If B TRIGGER level
readout indication is not within limits given in - peak column of table 7 for 2 mV setting,
perform b below.
(13) Repeat technique of (8) and (10) through (12) above for each remaining
VOLTS/DIV switch setting listed in
table 7.
Test
Test instrument trigger level readout indications
instrument
- Peak
VOLTS/DIV
+ Peak
switch
output
settings
settings
Min
Max
Min
Max
2
mV
10 mV
8.5 mV
11.5 mV
-1.2 mV
1.2 mV
5
mV
20 mV
17.3 mV
22.7 mV
-2.1 mV
2.1 mV
10
mV
50 mV
44.5 mV
55.5 mV
-4.0 mV
4.0 mV
20
mV
0.1 V
89 mV
111 mV
-7.5 mV
7.5 mV
50
mV
0.2 V
177 mV
223 mV
-17 mV
17 mV
100
mV
0.5 V
449 mV
551 mV
-36 mV
36 mV
10