Table14. A SWP Cursor Timing
Table 15. Delta Time Display

Calibration Procedure For Oscilloscope Types 2465A And 2465 Page Navigation
  14    15    16    17    18  19  20    21    22    23    24  
img
. . . .
TB 9-6625-2294-24
Table14. A SWP Cursor Timing - Continued
Test instrument
Oscilloscope
Test instrument
t readout indications
A SWP SEC/DIV
calibrator MARKER
switch settings
output settings
Min
Max
s
S/D
s
s
100
100
793.0
807.0
s
S/D
s
s
200
200
1586.0
1614.0
s
S/D
s
s
500
500
3965
4035
1  ms
1  mS/D
7.930
ms
8.070
ms
2  ms
2  mS/D
15.860
ms
16.140
ms
5  ms
5  mS/D
39.65
ms
40.35
ms
10  ms
10  mS/D
79.30
ms
80.70
ms
20  ms
20  mS/D
158.60
ms
161.40
ms
50  ms
50  mS/D
396.5
ms
403.5
ms
100  ms
100  mS/D
793.0
ms
807.0
ms
200  ms
200  mS/D
1578
ms
1622
ms
500  ms
500  mS/D
3945
ms
4055
ms
(24) Set A SWP SEC/DIV switch to 10 ns and B SWP SEC/DIV switch to 5 ns
(knob out).
(25) Press corresponding pushbuttons for indications as listed in (a) through (e) below.
TRIGGER SOURCE to VERT CH 1.
(a)
t to off (DLY readout).
(b)
A/B TRIG for B TRIGGER.
(c)
TRIGGER MODE to RUN AFT DLY.
(d)
XI0 MAG on.
(e)
(26) Set oscilloscope calibrator MARKER output initially for 10 nS/D.
(27) Set VOLTS/DIV switch as required for a display of 3 to 6 divisions and adjust
REF OR DLY POS control for DLY 10.64 ns.
(28) Adjust TRIGGER LEVEL control as required for a stable display.
(29) Adjust horizontal POSITION control cw until the trace stops moving, then ccw
to position leading edge of the 2d time marker near graticule center.
(30) Set oscilloscope calibrator MARKER output for 5 nS/D.
(31) Press t pushbutton to obtain t display and push SEC/DIV knob in for B SWP
only.
(32) Initially adjust  control ccw until signal trace movement stops. Then adjust
control cw to superimpose the two time markers.
t indication will be within limits
specified in first row of table 15. If not, perform b below.
(33) Adjust  control cw to move signal trace 2 signal peaks to left and superimpose
the signal trace (will be near 0.00 nSec on TI display).  t indication will be within limits
specified in second row of table 15. If not, perform b below.
(34) Repeat technique of (33) above for remaining  t readout indications listed in
table 15. If t readout indications are not within limits specified in table 15, perform b below.
19