TB 9-6625-2294-24
Table14. A SWP Cursor Timing - Continued
Test instrument
Oscilloscope
Test instrument
t readout indications
A SWP SEC/DIV
calibrator MARKER
switch settings
output settings
Min
Max
s
S/D
s
s
100
100
793.0
807.0
s
S/D
s
s
200
200
1586.0
1614.0
s
S/D
s
s
500
500
3965
4035
1 ms
1 mS/D
7.930
ms
8.070
ms
2 ms
2 mS/D
15.860
ms
16.140
ms
5 ms
5 mS/D
39.65
ms
40.35
ms
10 ms
10 mS/D
79.30
ms
80.70
ms
20 ms
20 mS/D
158.60
ms
161.40
ms
50 ms
50 mS/D
396.5
ms
403.5
ms
100 ms
100 mS/D
793.0
ms
807.0
ms
200 ms
200 mS/D
1578
ms
1622
ms
500 ms
500 mS/D
3945
ms
4055
ms
(24) Set A SWP SEC/DIV switch to 10 ns and B SWP SEC/DIV switch to 5 ns
(knob out).
(25) Press corresponding pushbuttons for indications as listed in (a) through (e) below.
TRIGGER SOURCE to VERT CH 1.
(a)
t to off (DLY readout).
(b)
A/B TRIG for B TRIGGER.
(c)
TRIGGER MODE to RUN AFT DLY.
(d)
XI0 MAG on.
(e)
(26) Set oscilloscope calibrator MARKER output initially for 10 nS/D.
(27) Set VOLTS/DIV switch as required for a display of 3 to 6 divisions and adjust
REF OR DLY POS control for DLY 10.64 ns.
(28) Adjust TRIGGER LEVEL control as required for a stable display.
(29) Adjust horizontal POSITION control cw until the trace stops moving, then ccw
to position leading edge of the 2d time marker near graticule center.
(31) Press t pushbutton to obtain t display and push SEC/DIV knob in for B SWP
only.
(32) Initially adjust control ccw until signal trace movement stops. Then adjust
control cw to superimpose the two time markers.
t indication will be within limits
specified in first row of table 15. If not, perform b below.
(33) Adjust control cw to move signal trace 2 signal peaks to left and superimpose
the signal trace (will be near 0.00 nSec on TI display). t indication will be within limits
specified in second row of table 15. If not, perform b below.
(34) Repeat technique of (33) above for remaining t readout indications listed in
table 15. If t readout indications are not within limits specified in table 15, perform b below.
19