TB 9-6625-2294-35
Test instrument parameters
Performance specification
Range: 2 mV/div to 5 V/div
Vertical gain CH I and CH 2
Accuracy: 2% (1 MΩ input coupling)
Delta V cursors
Accuracy: (1.25% of reading + .03 div)
CH 3 and CH 4
Range: .1 and .5 V/div
Accuracy: 10%
Trigger level readout
Accuracy: Within [3% of setting + 3% of p-p signal
+ 0.2 division + 0.5 mV +(0.5 mV x probe
Trigger sensitivity CH I and CH 2
Dc coupled
0.35
division
Dc
to
50
MHz
1.0
division
at
300
MHz
1.5
division
at
500
MHz
Ac coupled
0.35
division
60 Hz
to
50
MHz
1.0
division
at
30
MHz
1.5
division
at
500
MHz
Horizontal timing
Range: 500 ms/div to 5 ns/div extended to 0.5 ns with X10 mag
A sweep
Accuracy: 500 and 200 ms (1.2% of time interval + 0.6% of FS)
100 ms to 5 ns (0.7% of time interval + 0.6% of FS)
X10 mag:(1.2% of time interval + 0.6% of FS)
B sweep
Range: 50 ms/div to 5 ns/div extended to 0.5 ns with X10 mag
Accuracy: 1.0% of time interval + 0.6% of FS
X10 mag:1.5% of time interval + 0.6% of FS
Accuracy: (0.5% of time interval + 0.3% of FS)
Delta time with cursors
X10 mag:(1% of time interval + 0.3% of FS)
(0.3% of time interval + 0.1% of FS + 200 ps)
Accuracy:
Delta time with sweep delay
(0.8% of time interval + 0.1% of FS + 200 ps)
X10 mag:
(0.8% of time interval + 0.3% of FS)
Accuracy:
Delta time with cursors on
(1.3% of time interval + 0.3% of FS)
X10 mag:
B sweep
(0.3% of delay setting + 0.6% of FS) +0 to -25 ns
Accuracy:
Delay sweep
DC to 350 MHz
Calibrator Output voltage
0.4V 1% into 1 MΩ load
tested below 50 kHz.
1Not