TB 9-6625-2294-35
(14) Press VERTICAL MODE CH 2 pushbutton off and CH 3 on.
(15) Press CH 3 VOLTS/DIV pushbutton for .1 V.
NOTE
press blue soft pushbutton located below CHAN 3 on
oscilloscope calibrator when SELECT CHANNEL is displayed.
(16) Ensure oscilloscope calibrator SOURCE/MEASURE CHAN 3 is connected to TI
CH 3 using a 50Ω feedthrough termination.
(17) Set oscilloscope calibrator LEVEL SINE frequency to 50 kHz and output
initally for 600 mV. Rotate knob located below EDIT FIELD pushbutton for 6 divisions of
display on TI.
divisions in amplitude or greater through out sweep; if not, perform b below.
(19) Press TRIGGER SOURCE lower pushbutton to CH 4.
(20) Press VERTICAL MODE CH 3 pushbutton off and CH 4 on.
(21) Press CH 4 VOLTS/DIV pushbutton for .1 V.
NOTE
press blue soft pushbutton located below CHAN 4 on
oscilloscope calibrator when SELECT CHANNEL is displayed.
(22) Ensure oscilloscope calibrator SOURCE/MEASURE CHAN 4 is connected to TI
CH 4 using a 50Ω feedthrough termination.
(23) Set oscilloscope calibrator LEVEL SINE frequency to 50 kHz and output
initally for 600 mV. Rotate knob located below EDIT FIELD pushbutton for 6 divisions of
display on TI.
divisions in amplitude or greater through out sweep, if not, perform b below.
NOTE
press blue soft pushbutton located below CHAN 1 on
oscilloscope calibrator when SELECT CHANNEL is displayed.
(25) Set oscilloscope calibrator output to SBY and remove all connections.
25