TB 9-6625-2295-35
Paragraph
Page
SECTION
Preliminary instructions .......................................
Equipment setup ....................................................
Power supply ..........................................................
DAC REF and input capacitance ..........................
CAL 01 horizontal ..................................................
CAL 02 vertical.......................................................
CAL 03 triggering ..................................................
CAL 06 vertical transient response ......................
CAL 09 parametric measurements .......................
Final procedure ......................................................
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
OPT46 and 2467B. The manufacturer's manual was used as the prime data source in
compiling these instructions. The equipment being calibrated will be referred to as the TI
(test instrument) throughout this bulletin.
a. Model Variations. Tektronix, Type 2465B OPT46 and 2467B OPT10 are equipped
with an IEEE bus.
b. Time and Technique. The time required for this calibration is approximately 6
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables the (R) follows the designated adjustments.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Range: 2 mV/div to 5 V/div
Vertical gain CH1 and CH2
Accuracy: 2% (1 MΩ input coupling)
3% (50 Ω input coupling)
∆V cursors
Accuracy: (1.25% of reading + .03 div)
CH 3 and CH 4
Range: .1 and .5 V/div
Accuracy: 10%