TB 9-6625-2296-24
Table 1. Calibration Description Continued.
Test instrument parameters
Performance specifications
FM Deviation:
Range
100 Hz to 25.0 kHz
Modulation Rates
30 Hz to 20 kHz
Accuracy
5% at 1.0 thru 20.0 kHz
10% at 100 Hz to <1.0 kHz and >20.0 kHz to 25.0 kHz
(sine wave only):
>6 kHz deviation (sine wave only)
<0.7% (700 Hz thru 1.1 kHz); 0.3 to 3 kHz post detection filter
<1.0% (30 Hz thru 10 kHz)
<2.0% (> 10.0 thru 20.0 kHz)
AM Deviation:
Range
750 kHz to 999.9999 MHz (1 to 90%)
Modulation Rates
100 Hz to 10 kHz
Accuracy
5% of setting (30% thru 90% mod)
10% of setting (30% thru 90% mod) fc ≤1 MHz
70%)
<0.7% (700 Hz thru 1.1 kHz); 0.3 to 3 kHz post detection filter
(sine wave only)
<1.5% (100 Hz thru 6.0 kHz)
<2.5% (> 6 thru 10.0 kHz)
<5.0% (0.75 thru 1.0 MHz)
POWER METER
Frequency Range
1.5 to 999.9999 MHz
Level Range
0.2 mW to 200 W (up to 30 MHz)
0.2 mW to 100 W (>30 to 200 MHz)
0.2 mW to 50 W (>200 to 999.9999 MHz)
Accuracy
10% (up to 400 MHz) 1 count
20% (beyond 400 MHz) 1 count
250 kHz to 999.9999 MHz
Frequency Range
-10 to 50 dBm (250 kHz to <200 MHz)
Level Range
-10 to 47 dBm (200 to 999.9999 MHz)
0.5 PPM
Accuracy
0 to 150 kHz
Counter Range
0 to 100 kHz
Meter Range
-10 to +50 dBm at T/R connector
Level Range
0.5 ppm
Accuracy
Range
10.0 Hz to 40.0 kHz
Accuracy
0.5 ppm
See footnotes at end of table.
3