Figure 2. Top View.
Distortion Meter

Calibration Procedure For Radio Test Set An/Grm-114B, An/Grm-122, And Ts-4317/Grm Page Navigation
  15    16    17    18    19  20  21    22    23    24    25  
img
. . .
TB 9-6625-2296-24
(d) DATA ENTRY 3 ....................
Wave Form
(e) DATA ENTRY 7 ....................
0 Lvl
(f) ESC (F6) .................................
Escape
(g) DATA ENTRY 2 ....................
AF Gen # 2 Setup
(h) DATA ENTRY 3 ....................
Wave Form
(i) DATA ENTRY 7 ....................
0 Lvl
ESC (F6) .................................
(j)
Escape
(k) DATA ENTRY 5 ....................
AF Output Setup
(l) DATA ENTRY 2 ....................
To Speaker On
(m) Ret (F5) ..................................
Return
(4) Connect multimeter INPUT HI and LO to TI AUDIO OUT. Set multimeter to
measure volts dc.
(5) Press FIELD SELECT  key to select LEVEL and press DATA ENTRY
ENTER key. Press DATA ENTRY 3 . 1 and ENTER keys.
(6) Press FIELD SELECT  key to select GEN 1 and press DATA ENTRY
ENTER key to set GEN 1 On.
(7) If multimeter does not indicate 0.000 0.01 V dc adjust A24R23 (fig. 3) as
close as possible.
Figure 3. Function Generator Assembly A24.
(8) Press DATA ENTRY ENTER key to set GEN 1 Off. Press FIELD SELECT 
key to select GEN 2 and press DATA ENTRY ENTER key to set GEN 2 On.
(9) If multimeter does not indicate 0.000 0.01 V dc adjust A24R22 (fig. 3) as close
as possible.
(10) Press DATA ENTRY ENTER key to set GEN 2 Off. Press FIELD SELECT 
key to select GEN 1 and press DATA ENTRY ENTER key to set GEN 1 On.
20