TB 9-6625-2296-24
(5) Press Func (F1) key, adjust DATA SCROLL knob to select DCV, and press
DATA ENTRY ENTER key.
(6) Press Range (F2) key, adjust DATA SCROLL knob to select TI range settings,
press DATA ENTRY ENTER key, and set calibrator for output settings listed in table 7.
TI multimeter indications will be within limits specified in table 7.
Test instrument
Test instrument
Calibrator
output settings
range settings
Min
Max
200
mV
190
mV
187.9
mV
192.1
mV
2.0
V
1.9
V
1.879
V
1.921
V
20
V
19
V
18.79
V
19.21
V
200
V
190
V
187.9
V
192.1
V
V1
980
V
959
V
1001
V
2000
1After
verification press RESET key on calibrator.
(7) Connect calibrator OUTPUT HI and LO to TI AMP and COM.
(8) Press Func (F1) key, adjust DATA SCROLL knob to select DCC, and press
DATA ENTRY ENTER key.
(9) Press Range (F2) key, adjust DATA SCROLL knob to select TI range settings,
press DATA ENTRY ENTER key, and set calibrator for output settings listed in table 8.
TI multimeter indications will be within limits specified in table 8.
NOTE
If current shunt is not supplied with TI, the 20 A check in
table 8 cannot be performed and a DA Label 163 should be
affixed stating such.
Test instrument
Test instrument
Calibrator
range settings
output settings
Min
Max
20 mA
18
mA
16.99 mA
19.01 mA
200
mA
180
mA
169.9
mA
190.1
mA
2
A
1.8
A
1.699 A
1.901 A
A1
20
5
A
3.9
A
6.1
A
1 Set
calibrator to STANDBY and disconnect from TI. Connect current shunt (if supplied
with TI) to TI DMM V/Ω and COM. Connect calibrator amplifier OUTPUT HI and LO to
current shunt inputs observing polarity. After verification set calibrator to STANDBY.
(10) Connect calibrator OUTPUT HI and LO to TI DMM V/Ω and COM.
(11) Press Func (F1) key, use DATA SCROLL controls to select Ohm, and press
DATA ENTRY ENTER key.
(12) Press RANGE (F2) key, use DATA SCROLL controls to select TI range
settings, press DATA ENTRY ENTER key, and set calibrator output for settings listed in
table 9. At each setting, use calibrator output adjustment controls to set calibrator control
26