TB 9-6625-2303-24
14. Plug-in Element 4410-11 Calibration
a. Performance Check
(1) Connect equipment as shown in figure 2 above.
(2) Insert plug-in element into TI and rotate until the arrow points in the direction
of the 50 Ω termination.
(3) Set TI range switch to 100 and allow equipment to warm-up for 5 minutes.
(5) Adjust signal generator and RF power amplifier amplitude controls for a TI
indication of 100 W. RF power meter will indicate within limits in table 9.
(6) Repeat technique of steps (4) and (5) above for remaining settings listed in
Test instrument
Test instrument
RF power meter indication (W)
Signal generator
Low pass
range switch
full scale
filter used
Min
Max
settings
reading (W)
100
100
95
105
30
30
28.5
31.5
10
10
9.5
10.5
125 M
TLC125-6EF1
3
3
2.85
3.15
1
1
0.95
1.05
0.3
0.3
0.285
0.315
0.1
0.1
0.095
0.105
0.1
0.1
0.095
0.105
0.3
0.3
0.285
0.315
1
1
0.95
1.05
85 M
TLC125-6EF1
3
3
2.85
3.15
10
10
9.5
10.5
30
30
28.5
31.5
100
100
95
105
100
100
95
105
30
30
28.5
31.5
10
10
9.5
10.5
50 M
TLC75-6EF1
3
3
2.85
3.15
1
1
0.95
1.05
0.3
0.3
0.285
0.315
0.1
0.1
0.095
0.105
(7) Reduce outputs to minimum and remove plug-in element from TI.
b. Adjustments. No adjustments can be made.