TB 9-6625-2315-35
(6) Press MENU < or > key until F: CAL MENU is displayed.
(7) Press RANGE/DIGITS ∨ key until 1: UNSECURED is displayed.
(8) Press MENU > key until 2: CALIBRATE is displayed.
(9) Press RANGE/DIGITS ∨ key until PARAMETERS is displayed.
NOTE
The calibration parameter initial display will be 0.000,000 HZ.
HZ will have to be changed to 00.000,00 kHz in order to enter
50.000,00 kHz in (10) below.
(10) Press MENU < or > and RANGE/DIGITS ∧ or ∨ keys until 50.000,00 kHz is
displayed.
(11) Press RANGE/DIGITS Auto/Man key. Adjustment takes approximately 20
seconds to complete then TI will return to measurement mode.
from TI.
g. Press TI keys as listed in (1) through (6) below:
(1) Shift then MENU < to display A: MEAS MENU.
(2) MENU > until F: CAL MENU is displayed.
(3) RANGE/DIGITS ∨ until 1: UNSECURED is displayed.
(4) RANGE/DIGITS ∨ until ^000000 CODE is displayed.
(5) MENU < or > and RANGE ∧ and ∨ until 034401 is displayed.
(6) RANGE/DIGITS Auto/ Man. TI is now in secure mode.
a
Deenergize and disconnect all equipment
b. Annotate and affix DA label/form in accordance with TB 750-25.