TB 9-6625-2325-24
Table 1. Calibration Description - Continued
Test instrument
Performance Specifications
parameters
Linear analog output
Range: 0 to 2 V dc
Accuracy: 1%
(model 8920A only)
Ac + dc voltage
Range: 0 to 700 V
Accuracy: Add to ac specifications:
Above 2 mV (10 digits or 0.5 dB)
Below 2 mV (100 digits or 5 dB)
Use 50 Hz to 200 kHz specifications for dc only.
.
dB display mode
Range: 12 ranges: 50, 75, 93, 110, 124, 135, 150, 300, 600, 900, 1000, and 1200
dB resolution: 0.01 dB
1Maximum
V Hz product is 1 x 108.
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-286, AN/GSM-287 or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this