TB 9-6625-2325-35
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Dc voltage:
John Fluke, Model 5700A/CT
Range: 1 mV to 2.5 V
(p/o MIS-35947); w/
Accuracy: 1 mV, 0.375%
amplifier, John Fluke, Model
1 V, 4%
5725A (5725A)
0.25 and 2.5 V, 1.12%
Ac voltage:
Range: 1 mV to 600 V
Accuracy (%)
Voltage
10
Hz
1.9
mV to
190
V
1.25
40
Hz
1.9
mV
0.75
19
mV
0.5
190
mV to
600
V
0.25
500
Hz
1.9
mV
0.5
100
mV to
100
V
0.125
1
kHz
1
mV
0.5
10
mV
0.25
100
mV to
3
V
0.125
5
kHz
1.9
mV
0.5
19
mV
0.25
190
mV to
190
V
0.125
600
V
0.125
40
kHz
600
V
0.125
50
kHz
100
mV to
100
V
0.125
100
kHz
19
mV
0.25
190
mV to
190
V
0.125
190
kHz
1
mV
0.5
990
kHz
1
mV
0.75
2
MHz
1
mV
1
190
kHz
10
mV
0.25
990
kHz
10
mV
0.5
5
MHz
10
mV
1
20
MHz
10
mV
1.25
190
kHz
100
mV to
3
V
0.125
990
kHz
100
mV to
3
V
0.175
5
MHz
100
mV to
3
V
0.75
20
MHz
100
mV to
3
V
1.25
Range: -15.2 to +15.1 V dc
John Fluke, Model 8840A/AF
Accuracy: 0.17%
(AN/GSM-64D)
4