TB 9-6625-2328-35
(4) Press MATH Scale & Offset key until CODE:
0 is displayed.
(5) Press ⇐ key until CODE: 00000 is displayed.
(6) Press ⇒ , ⇓ , ⇐ , ⇑ keys until CODE: 53131 is displayed then press Enter key
and verify CAL UNSECURE is displayed.
(7) Press MATH Scale & Offset key until CAL: OFFS 1 ? is displayed.
(8) Press Enter key. When THEN ENTER is displayed press Enter key.
(9) Verify CALIBRATING is displayed followed by OFFS 1 PASS.
(10) Press ⇒ key until CAL: OFFS 2 ? is displayed.
(11) Press Enter key. When THEN ENTER is displayed press Enter key.
(12) Verify CALIBRATING is displayed followed by OFFS 2 PASS.
(13) Connect calibrator OUTPUT HI and LO to TI CHANNEL 1 input.
(14) Set calibrator for a 5 V dc output.
(15) Press ⇒ key until CAL: GAIN 1 ? is displayed.
(16) Press Enter key. When THEN ENTER is displayed, press Enter key.
(17) Verify CALIBRATING is displayed followed by GAIN 1 PASS.
(18) Connect calibrator OUTPUT HI and LO to TI CHANNEL 2 input.
(19) Set calibrator for a 5 V dc output.
(20) Press ⇒ key until CAL: GAIN 2 ? is displayed.
(21) Press Enter key. When THEN ENTER is displayed, press Enter key.
(22) Verify CALIBRATING is displayed followed by GAIN 2 PASS.
(23) Press MATH Scale & Offset key until CODE:
0 is displayed.
(24) Press ⇐ key until CODE: 00000 is displayed.
(25) Press ⇒ , ⇓ , ⇐ , ⇑ keys until CODE: 53131 is displayed then press Enter key
and verify CAL SECURE is displayed.
(26) Disconnect calibrator from TI.
a. Performance Check
(1) Press and release TI POWER ON/STBY pushbutton to STBY.
(2) Press TI POWER ON/STBY pushbutton to ON.