Time Base Stability
Table 3. Input A and B Sensitivity.

Calibration Procedure For Electronic Counter, Model Pm6681/656 Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . . . .
TB 9-6625-2331-24
Front
Figure 1. Top View.
(9) Adjust autotransformer output to 125 V and allow 15 minutes for stabilization.
Verify that TI oscillator drift is less than 5 parts in 10-10.
(10)Adjust autotransformer output to 115 V.
b. Adjustments. No further adjustments can be made.
9. Sensitivity
a. Performance Check
(1) Press TI LOCAL/PRESET key.
(2) Press both INPUT A 50Ω/1MΩ and INPUT B 50Ω/1MΩ key to select 50Ω.
(3) Connect signal generator RF output to TI INPUT A.
(4) Set signal generator to settings in first row of table 3 with RF output off.
(5) Set the signal generator RF output to on. If TI does not indicate a stable display
of the applied frequency as listed in table 3, perform b below.
(6) Repeat steps (4) and (5) above for remaining settings listed in table 3.