TB 9-6625-2333-35
11. Input Attenuator Accuracy
a. Performance Check
Figure 1. Input attenuator accuracy (50 MHz) -
equipment setup.
(2) Press synthesizer/level generator keys to values as listed in (a) through
(c) below:
(b) AMPLITUDE to -50 dBm.
(c) AMPTD INCR to 10 dB.
(3) Press TI keys and enter values using DATA keys as listed in (a)
through (i) below:
(a)
INSTRUMENT STATE PRESET.
(b)
INSTRUMENT STATE CAL.
(c)
[REALIGN LO & IF] wait for ADJUST STATUS messages to disappear.
(d)
(e)
SPAN then [ZERO SPAN].
(f)
AMPLITUDE then [REF LVL] to -70 dBm.
(g)
[LOG dB/DIV] to 1 dB.
(h)
CONTROL BW then [RES BW] to 3 kHz.
(i)
[VIDEO BW] to 1 Hz.
(4) Set attenuator to 0 dB.
(5) Adjust attenuator to position signal peak 2 to 3 divisions below TI reference level.