TB 9-6625-2334-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Universal Counter, Hewlett-Packard, Model 5328B( ). The manufacturer's
manual was used as the prime data sources in compiling these instructions. The
equipment being calibrated will be referred to as the TI (test instrument) throughout this
bulletin.
a. Model Variations: Model 5328B may have the following options:
010
021
031
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Time base stability (Option 010):
Frequency: 10 MHz
Aging rate (after 24 hr warm-up): <5 X 10-10/day
Line voltage stability (for 10% variation): <5 X 10-9
Channel A and B sensitivity:
25 mV rms:
Dc to 40 MHz (dc coupled)
20 Hz to 40 MHz (ac coupled)
50 mV rms:
40 to 100 MHz
Channel C sensitivity (Option 031):
20 mV rms:
90 to 1300 MHz
Range: 10, 100, 1000 V dc and autoranging
Accuracy:
10, 100 V ranges: 0.03% of reading, 0.004% of range
1000 V range: 0.087% of reading, 0.004% of range