TB 9-6625-2335-35
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set, AN/GSM-286. Alternate items may be used by the calibrating activity.
The items selected must be verified to perform satisfactorily prior to use and must bear
evidence of current calibration. The equipment must meet or exceed the minimum use
5. Accessories Required. The
accessories required for this
calibration are
issued as
indicated in 4 above. Common usage accessories are not listed in this calibration
procedure.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Accuracy (%)
CALIBRATOR
Dc voltage
John Fluke, Model 5720A
30 mV
0.004
(p/o MIS-35947); w/power
300 mV
0.001
amplifier, John Fluke Model
3 V
0.0009
5725A (5725A)
30 V
0.0012
300 V
0.0014
Ac voltage range: 30 mV to 300 V
Frequency: 20 Hz to 1 MHz
Accuracy: (%)
Voltage
0.02
0.05
1
10
90
250
1000
30 mV
0.16
0.073
0.053
0.56
0.19
0.88
3.1
300 mV
0.16
0.073
0.053
0.56
0.19
0.88
3.1
3V
0.16
0.073
0.053
0.56
0.19
0.88
3.1
30 V 1
0.16
0.073
0.053
0.56
0.19
0.88
---
300 V 2
---
0.087
0.073
---
0.31
---
---
Accuracy (%)
≤837 Ω
0
10
0.01
100
0.05
1k
0.006
10 k
0.002
100 k
0.001
1M
0.002
10 M
0.01
FUNCTION/ARBITRARY
Function: Sine wave
Hewlett-Packard, Model
Amplitude: 1 V rms
33250A (MIS-45853)
Frequency
Accuracy (%)
20 Hz
0.0125
1 MHz
0.0025