TB 9-6625-2335-35
(5) Press RANGE ↑ or ↓ key to select ranges listed in table 9. If TI indications
are not within limits specified, perform b(1), (2) and (4) below.
Test instrument
Indications (≤)
Ranges
30 Ω
00.00335 Ω
300 Ω
000.0035 Ω
3 kΩ
0.000007 kΩ
30 kΩ
00.00007 kΩ
300 kΩ
000.0008 kΩ
3 MΩ
0.000014 MΩ
30 MΩ
00.00083 MΩ
(6) Press TI RANGE ↑ or ↓ key to select ranges and set calibrator output for
settings listed in table 10. At each output setting use calibrator output adjustment
controls to set calibrator control display Reading equal to TI indication. If calibrator
control display Error indications are not within limits specified, perform b(1), (3) and
(4) below.
Calibrator
Control display
Test instrument
Error indications
(%)
ranges
Output settings
30 Ω
10 Ω
0.041
300 Ω
100 Ω
0.009
3 kΩ
1 kΩ
0.0057
30 kΩ
10 kΩ
0.0057
300 kΩ
100 kΩ
0.0058
3 MΩ
1 MΩ
0.0079
30 MΩ
10 MΩ
0.0483
b. Software Adjustments
(1) Press blue shift key then CONFIGURATION C/OFFSET COMP key. TI
will display CAL.
(2) Enter 0,3457 using MATH keys.
(3) Enter calibrator output display indication rounded to TI digits of resolution
,3457 using MATH keys.
(4) Press ENT key.
TI will display CALIBRATING while the adjustment
routine is running.
a. Performance Check
14