TB 9-6625-2337-35
(6) Repeat technique of (4) and (5) above using remaining setting and limits listed
Table 27. Suppression-Gate Output Characteristics
TI
Oscilloscope
Test description
Settings
Measurement type
Minimum
Maximum
SUPPR OUT RT
----------------------------------------
Rise
--------------------
400 ns
SUPPR OUT FT
----------------------------------------
Fall
--------------------
400 ns
24 s
25 s
SUPPR OUT 25 s
----------------------------------------
Width
PW
SUPPR OUT 20 V
----------------------------------------
Top
15 V
25 V
Lvl
22 s
23 s
SUPPR OUT
----------------------------------------
Delta time
TIMING
Channel 1 1st
pulse to channel 2
2d pulse
0.1 s
3 s
SUPPR OUT 2 s
Use arrows to highlight
Width
GENERAL MENUS and press
PW
ENTR
15, ENTR (SUPPRESSION)
Use the arrow keys to highlight
SUPPRESSION WIDTH:
2, ENTR
219 s
221 s
SUPPR OUT
Use the arrow keys to highlight
Width
220 s PW
SUPPRESSION WIDTH:
220, ENTR
Press FUNC and ↑ keys
15 s
25 s
M4 SYNC DELAY
Delta time
Channel 1 1st
14, ENTR (PRF/PRI)
pulse to channel 2
Use the arrow keys to highlight
4th pulse
PRF:
1000, ENTR (1000 PPS)
Press FUNC and ↑ keys
2, ENTR (CHALLENGES)
Use the arrow keys to highlight
M2:
ENTR (OFF)
Use the arrow keys to highlight
M4:
ENTR (SYNC)
(7) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect TI TRIGGER 0 OUT to oscilloscope channel 1 using a 75 Ω
feedthrough termination.
(2) Connect TI TRIGGERS M4 PRE OUT to the oscilloscope channel 2 input using
a 93 Ω feedthrough termination.
(3) Press TI FUNC, ENTR keys and wait for TI to complete reset.