(16) Disconnect oscilloscope Vertical 1 input from TI B output.
(17) Connect oscilloscope Vertical 1 input to TI +AB output.
(18) Press and hold down the TI BSP key then press and release the TI GPIB key
until TI display indicates Amplitude AB
###.
(19) Release the BSP key.
(20) Use the TI left/right (4, 6) and the up/down (8, 2) cursor control keys to adjust
(21) Disconnect oscilloscope Vertical 1 input from TI +AB output.
(21) Connect oscilloscope Vertical 1 input to TI C output.
(23) Press and hold down the TI BSP key then press and release the TI GPIB key
until TI display indicates Amplitude C
###.
(24) Release the TI BSP key.
(25) Use the TI left/right (4, 6) and the up/down (8, 2) cursor control keys to adjust
(26) Disconnect oscilloscope Vertical 1 input from TI C output.
(27) Connect oscilloscope Vertical 1 input to TI D output.
(28) Press and hold down the TI BSP key then press and release the TI GPIB key
until TI display indicates Amplitude D
###.
(29) Release the TI BSP key.
(30) Use the TI left/right (4, 6) and the up/down (8, 2) cursor control keys to adjust
(31) Disconnect oscilloscope Vertical 1 input from TI D output.
(32) Connect oscilloscope Vertical 1 input to TI +CD output.
(33) Press and hold down the TI BSP key then press and release the TI GPIB key
until TI display indicates Amplitude +CD
###.
(33) Release the TI BSP key.
(35) Use the TI left/right (4, 6) and the up/down (8, 2) cursor control keys to adjust
a. Performance Check
(1) Connect TI T0 output to oscilloscope Vertical 1 input.
(2) Press TI keys as listed in (a) through (f) below:
(a)
Press RECALL, 0, EXC.
(b)
4 (cursor left), 4.
(c)
TRIG, NUM/REM, 100000, EXC.
DELAY, DELAY, NUM/REM, 5, EXP, -, 6 (cursor right), EXC.
(d)
DELAY, DELAY, NUM/REM, 5, EXP, -, 6, EXC.
(e)