TB 9-6625-2342-35
n. Press A/B TRIG button.
o. Select TRIGGER MODE and set TRIG AFTER to ON.
p. Press A/B TRIG button.
q. Select STORAGE ACQUIRE and set REPET ON: OFF pushbutton to ON.
r. Press AVG switch until 16 appears above AVG.
s. Press ENVELOPE switch until 16 appears above ENVELOPE.
t. Set NORMAL to on.
u. Select PRGM to display SEQUENCER menu.
v. Press SAVE in main menu to display SAVE SEQUENCER menu.
NOTE
Saving FPNL sequence will cause an error if a FPNL sequence
is already stored. Clear the previous sequence by selecting
PRGM, pressing DELETE, pressing the up/down arrows to
select FPNL, then press DELETE.
w. Use arrows under ROLL-CHARS to create a front panel setup label (use FPNL for
the label).
x. Use the up/down arrow switches to select first letter of label.
y. When first letter is displayed press CURSOR <>.
z. Repeat technique of x and y above until FPNL is spelled out, then press menu
SAVE switch.
aa. Depressing menu switch saves the label for the sequence and displays SETUP
CONTROLS PUSH PRGM to CONTINUE. Select TRIGGER MODE and set AUTO
LEVEL.
ab. Press PRGM button then push SAVE SEQ.
a. Press PRGM button then press RECALL button in main menu.
b. Use arrow labeled switches to underline FPNL.
c. Press RECALL button.
d. Select VERTICAL MODE and set CH 2 to off.
e. Select CH 1 COUPLING/INVERT and set 50 Ω ON/OFF to OFF.
f. Select CH 2 COUPLING/INVERT and set 50 Ω ON/OFF to OFF.
h. Set CH 1 VOLTS/DIV to 100 mV.
i. Connect TI CALIBRATOR output to CH 1 input.