TB 9-6625-2343-40
(11) Adjust A6R65 ALC REF (fig. 3) for a display between 2.0970 and 2.1030 (R).
(12) Disconnect TI sensor module from TI CALIBRATION RF POWER OUTPUT.
(13) Connect TI sensor module to AM/FM test source FM OUTPUT.
(14) Connect calibrator WIDEBAND output to AM/FM test source AUDIO INPUT
using cable and termination supplied with calibrator.
(15) Set calibrator for a wideband output of 100 kHz at an amplitude of 500 mV.
R65 ALC REF
ALC REFERENCE
A6
(16) Set AM/FM test source TEST MODE switch to FM.
(17) Press TI keys as listed in (a) through (c) below:
(a) BLUE.
(b) INSTR PRESET.
(c) HP FILTER 50 Hz.
(18) Adjust AM/FM test source CARRIER FREQUENCY TUNE for a TI indication
of approximately 400 MHz.
13