Amplitude Modulation - Continued
Figure 4. A18 board.

Calibration Procedure For Measuring Receiver Model 8902A Page Navigation
  7    8    9    10    11  12  13    14    15    16    17  
img
. . .
TB 9-6625-2343-40
Figure 2. A2 board.
(11) Adjust A6R65 ALC REF (fig. 3) for a display between 2.0970 and 2.1030 (R).
(12) Disconnect TI sensor module from TI CALIBRATION RF POWER OUTPUT.
(13) Connect TI sensor module to AM/FM test source FM OUTPUT.
(14) Connect calibrator WIDEBAND output to AM/FM test source AUDIO INPUT
using cable and termination supplied with calibrator.
(15) Set calibrator for a wideband output of 100 kHz at an amplitude of 500 mV.
R65 ALC REF
ALC REFERENCE
A6
Figure 3. A6 board.
(16) Set AM/FM test source TEST MODE switch to FM.
(17) Press TI keys as listed in (a) through (c) below:
(a) BLUE.
(b) INSTR PRESET.
(c) HP FILTER 50 Hz.
(18) Adjust AM/FM test source CARRIER FREQUENCY TUNE for a TI indication
of approximately 400 MHz.
13