TB 9-6625-2343-50
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
AM/FM TEST SOURCE
Range: AM; 0 to 95%
Hewlett-Packard, Model 11715A
Carrier; 12.5 MHz
(11715A)
Rate; 10 Hz to 10 MHz
ATTENUATOR FIXED
Range: 6 dB
Weinschel, Model 9918-6dB
Frequency: 2.5 MHz to 1.3 GHz
AUDIO ANALYZER
Output: Freq; 50 Hz to 40 kHz
Boonton, Model 1121 (1121)
Ampl; 100 mV to 3 V
Output Accuracy: 1%
Measurement: Distn; <0.025%
Capability
CALIBRATOR
Frequency range: 20 Hz to 30 MHz
Fluke, Model 5720A (5700A/EP)
Output range: 1 mV to 3VAC
(p/o MIS-35947)
Range: 0 to 50 VDC
Hewlett-Packard, Model 3458A
Accuracy: 0.015%
(3458A)
MEASURING SYSTEM
Power output: 1mW @ 50 MHz
Tegam, Model IIA; consisting of
Accuracy: 0.3% (0.7%)
1806 dual type 4 power meter;
M1111 thermistor mount
Range: Vert; 0.02V/div
(OS-303/G)
Horiz; 10 s/div
PISTON ATTENUATOR
Range: 20 to 80 dB
Weinschel, Model PA-2 (PA-2)
Accuracy: Test report values
RANGE CALIBRATOR
Range: 10 W to 100 mW
Hewlett-Packard, Model 11683A
Accuracy: 0.025%
(11683A)
(SG-1207/U)
Output level: +19 dBm to 500 MHz
+13 dBm to 1100 MHz
Modulation Capability: AM and FM
AM; 0 to 95%
FM Range; 0 to 400 kHz pk
deviation
Accuracy: AM; 10%
FM; 10%
(SG-1207/U)
Output level: +19 dBm to 500 MHz
+13 dBm to 1100 MHz
Modulation Capability: AM and FM
AM; 0 to 95%
FM Range; 0 to 400 kHz pk
deviation
Accuracy: AM; 10%
FM; 10%
CHANGE 1