TB 9-6625-2343-50
within limits specified in first row of table 25.
(27) Repeat technique of (26) above for frequency steps listed in table 25 below. TI
will display within limits specified in first row of table 25.
Signal generator No. 1
Test instrument
output
Indications
frequency
(dB)
(MHz)
Min
Max
-0.05
29.999
0.05
29.998
0.10
-0.10
-0.15
29.997
0.15
-0.20
29.996
0.20
29.995
0.25
-0.25
(28) Set signal generator No. 1 for output frequency to 30 MHz.
(29) Step signal generator No. 1 output frequency up by 1 kHz. TI will display
within limits specified in first row of table 26.
(30) Repeat technique of (29) above for frequency steps listed in table 26 below. TI
will display within limits specified in first row of table 26.
Signal generator No. 1
Test instrument
output
Indications
frequency
(dB)
(MHz)
Min
Max
30.001
0.05
-0.05
30.002
-0.10
0.10
-0.15
30.003
0.15
30.004
-0.20
0.20
30.005
0.25
-0.25
(31) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. None
a. Performance Check
(1) Press TI keys as listed in (a) and (b) below:
(a) BLUE.
(b) INSTR PRESET.
(2) Connect signal generator No. 1 RF OUTPUT to TI sensor module.
(3) Set signal generator No. 1 for an output frequency of 455 kHz at an output level of 3 dBm.
(4) Press TI keys as listed in (a) through (h) below:
(a)
DATA 100.
(b)
DATA MHz.
(c)
DATA 24.0.
(d)
DATA SPCL.