TB 9-6625-2343-50
(13) Move positive lead from A13TP12 to A13TP11 (fig. 21) and adjust A13R12 (fig.
21) until
digital multimeter indicates within
0.05 V of
threshold value. (R)
(14) Disconnect equipment setup.
b. Adjustments. Adjust A10R24 (fig. 20) for a
digital multimeter indication of 15 V
.01 V dc. (R)
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.