TB 9-6625-2347-35
(3) Connect digital multimeter low lead to TI TP501 (fig. 1) and multimeter hi lead
to TI TP500 (fig. 1). If digital multimeter does not indicate between -8.56 and -8.64 V dc,
perform b (2) below.
b. Adjustments.
(1) Adjust R952 (fig. 1) for a digital multimeter indication of 4.3 V dc (R).
(2) Adjust R946 (fig. 1) for a digital multimeter indication of -8.60 V dc (R).
a. Deenergize and disconnect all equipment.
b. Annotate and affix label/form in accordance with TB 750-25.
26