TB 9-6625-2348-35
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals for this TI.
d. Unless otherwise specified, all control and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Remove protective cover as needed to make adjustments.
Replace cover after
completing the adjustments.
b. Set function switch to V
.
a. Performance Check
(1) Connect function/arbitrary waveform generator to TI V Ω
and COM inputs.
(2) Press TI Hz button.
frequency of 19.000 kHz. TI will indicate within limits specified in first row of table 3.
(4) Repeat technique of (3) above for settings and indications listed in table 3. TI
will indicate within limits specified in table 3.
Function/arbitrary
Test instrument
waveform generator output
indication (kHz)
Amplitude (rms)
Min
Max
150 mV
19.000
18.998
19.002
150 mV
190.00
189.98
190.02
equipment setup.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect calibrator to TI V Ω
and COM inputs.
(2) Set function switch to mV
.
(3) Set calibrator output for 390.0 mV. If TI does not indicate within limits
specified for appropriate model in first row of table 3, perform b below.
6
CHANGE 3