TB 9-6625-2351-35
Synthesizer/level
Test instrument
AMPLITUDE [REF LVL]
∆MKR indications
generator amplitude
settings
settings
(dB)
(dBm)
(dBm)
Min
Max
-10
-20
-21
-19
-20
-30
-31
-29
-30
-40
-41
-39
-40
-50
-51
-49
-50
-60
-61
-59
-60
-70
-71
-69
-70
-80
-81
-79
a. Performance Check
(1) Connect signal generator RF Output to TI INPUT 50 Ω.
(3) Press TI keys and enter values using DATA keys listed in (a) through (v) below:
(a)
INSTRUMENT STATE PRESET.
(b)
(c)
SPAN then [SPAN WIDTH] to 1 MHz.
(d)
AMPLITUDE then [REF LEVEL] to -5 dBm.
(e)
[LOG dB/DIV] to 1 dB.
(f)
CONTROL BW then [RES BW] to 3 kHz.
(g)
MARKER PEAK SEARCH.
(h)
[SIG TRK ON].
(i)
SPAN then [SPAN WIDTH] to 10 kHz.
(j)
CONTROL BW then [RES BW] to 1 kHz.
(k)
MARKER ON.
(l)
[SIG TRK OFF].
MARKER PEAK SEARCH.
(m)
MARKER MKR→.
(n)
[MARKER→ CF].
(o)
[MARKER→ REF LVL].
(p)
(q)
MARKER OFF.
(r)
CONTROL TRIG.
(s)
[SINGLE].
(t)
[SINGLE] wait for completion of sweep.
(u)
MARKER PEAK SEARCH.
18