TB 9-6625-2352-35
Table 4a. TS-352/U Specifications
Test instrument parameters
Performance specifications
Dc voltage
0 to 1000 V at 1000 Ω per V
Range:
Accuracy: 3% of FS
0 to 1000 V at 20,000 Ω per V
Range:
Accuracy: 4% of FS
0 to 5000 V at 20,000 Ω per V
Range:
Accuracy: 6% of FS
Ac voltage
0 to 500 V at 1000 Ω per V
Range:
Accuracy: 4% at 400 Hz, 7% at 10 kHz of FS
1000 V at 1000 Ω per V
Range:
Accuracy: 5% at 400 Hz, 8% at 10 kHz of FS
0 to 10,000,000 Ω
Range:
Accuracy: 3% (percent values are in terms of meter arc
length, not of Ω indicated by meter
Dc current
Range:
0 to 10 A
Accuracy: 3% of FS
Table 4b. TS-352/U Dc Volts
Dc Voltage
Test instrument
Calibrator output
Test instrument
Dc volts
Indication
Initial
Limits
range
(V)
output (V)
Min
Max
Adjustments
2.51
2.5
2.5
2.40
2.60
10
10
10
9.60
10.40
10
6
6
5.60
6.40
10
2
2
1.60
2.40
None
50
50
50
48.0
52.0
250
250
250
240.0
260.0
500
500
500
480.0
520.0
See footnote at the end of the table.
8