TB 9-6625-2353-35
(2) Short TI leads. If TI display does not indicate between -00.5 and +00.5 perform
b (1) below.
(3) Connect TI V-Ω input to calibrator OUTPUT HI and TI COM input to
calibrator OUTPUT LO. Set calibrator for an output amplitude of 190 mV at an
output frequency of 1 kHz. If TI does not indicate within limits specified in first row of
(4) Repeat technique of (1) (b) and (3) above, using the settings and indications
Calibrator
Test instrument
Output
Output
Indication limits
amplitude
frequency
Range
Min
Max
190
mV
1 kHz
200 mV
188.6
mV
191.5
mV
190
mV
30 Hz
200 mV
186.7
mV
193.4
mV
190
mV
100 Hz
200 mV
188.6
mV
191.5
mV
190
mV
5 kHz
200 mV
180.0
mV
200.0
mV
1
1.90 V
5 kHz
2V
1.800 V
2.000
V
1.90 V
100 Hz
2V
1.886 V
1.915
V
1.90 V
30 Hz
2V
1.867 V
1.934
V
19.00 V
30 Hz
20 V
18.67 V
19.34
V
19.00 V
100 Hz
20 V
18.86 V
19.15
V
19.00 V
5 kHz
20 V
18.00 V
20.00
V
190.0 V
1 kHz
200 V
188.6
V
191.5
V
190.0 V
100 Hz
200 V
188.6
V
191.5
V
190.0 V
30 Hz
200 V
186.7
V
193.4
V
750
V
40 Hz
750 V
734
V
766
V
750
V
400 Hz
750 V
741
V
759
V
1
If TI does not indicate within limits specified perform b (3) below.
b. Adjustments.
(1) Short TI leads and adjust R206 (fig. 1) for a TI indication of 0.00.
(2) Set calibrator for an output amplitude of 190.0 mV at a frequency of 1 kHz and
adjust R203 (fig. 1) for a TI indication of 190.0 mV (R).
(3) Set TI to 2 V range and set calibrator for an output amplitude of 1.900 V at a
frequency of 5 kHz and adjust C101 (fig. 1) for a TI indication of 1.900 V (R).
a. Performance Check
(1) Position TI controls as listed in (a) through (c) below:
(a) AC DC pushbutton to DC (out).
(b) mA pushbutton to in position.
(c) 200 mV, 200 A, 200 Ω pushbutton to in position.
(2) Connect TI mA input to calibrator OUTPUT HI and TI COM input to calibrator
OUTPUT LO.
PIN: 081248-001