TB 9-6625-2354-35
Table 5a. Simpson, Model 260-6 Specifications
Test instrument parameter
Performance specification
Dc voltage
Range: 0 to 1000 V
2% of FS
Accuracy:
Ac voltage
Range: 0 to 1000 V
3% of FS
Accuracy:
Range: 0 to 20 MΩ
2.5 of arc for RX1 range, 2 of arc for
Accuracy:
all other ranges
Dc current
Range: 0 to 10 A
1.5% of FS for 50 A range, 2% of FS for
Accuracy:
all other ranges
Table 5b. Simpson, Model 260-6 Dc Voltage
Dc Voltage
Test instrument1
Calibrator
Test instrument
Indication
Initial output
err indication
Dc volts range
(V)
(V)
(%)
Adjustments
1
1
1
2
2
2.5
2.5
2.5
2
10
10
10
2
50
50
50
2
None
250
250
250
2
500
500
500
2
10003
1000
1000
2
TI must be calibrated in horizontal position.
1
Connect positive lead to TI +1V DC input. After 1V check is complete, reduce output to minimum and move positive lead to
2
TI + input.
3 Reduce output to minimum and move positive lead from TI + input to TI 1000V DC/AC input. After 1000V check is
complete, reduce output to minimum and move positive lead to TI + input.
Table 5c. Simpson, Model 260-6 Ac Voltage
Ac Voltage
Test instrument
Calibrator
Test instrument
1
Ac volts
Indication
Initial output
err indication
Adjustments
range
(V)
(V)
(Hz)
(%)
2.5
2.5
2.5
60
3
R25
10
10
10
60
3
----
50
50
50
60
3
----
250
250
250
60
3
R22
500
500
500
60
3
----
1000
1000
1000
60
3
----
2
TI must be calibrated in horizontal position.
1
Reduce output to minimum and move positive lead from TI + input to TI 1000V DC/AC input. After 1000V check is
2
complete, reduce output to minimum and move positive lead to TI + input.
9