TB 9-6625-2354-35
Table 7b. Simpson, Model 260-6P Dc Voltage
Dc Voltage
Test instrument
Calibrator
Test instrument
1
Indication
Initial output
err indication
Dc volts range
(V)
(V)
(%)
Adjustments
12
1
1
2
2.5
2.5
2.5
2
10
10
10
2
50
50
50
2
None
250
250
250
2
500
500
500
2
1000
1000
1000
2
3
TI must be calibrated in horizontal position.
1
Connect positive lead to TI +1V DC input. After 1V check is complete, reduce output to minimum and move positive lead to
2
TI + input.
3 Reduce output to minimum and move positive lead from TI + input to TI 1000V DC/AC input. After 1000V check is
complete, reduce output to minimum and move positive lead to TI + input.
Table 7c. Simpson, Model 260-6P Ac Voltage
Ac Voltage
Test instrument1
Calibrator
Test instrument
Ac volts
Indication
Initial output
err indication
Adjustments
range
(V)
(V)
(Hz)
(%)
2.5
2.5
2.5
60
3
R25
10
10
10
60
3
----
50
50
50
60
3
----
250
250
250
60
3
R22
500
500
500
60
3
----
1000
1000
1000
60
3
----
2
TI must be calibrated in horizontal position.
1
Reduce output to minimum and move positive lead from TI + input to TI 1000V DC/AC input.
2
Resistance
(Ω)
Test instrument
Test instrument
1
Indications
Resistance
ohms scale
range
(Ω)
Min
Max
Adjustments
2
RX1
12
10.7
13.5
RX100
12
1110
1300
None
RX10,000
12
111,000
130,000
TI must be calibrated in horizontal position.
1
2 Short
leads and adjust OHMS ADJ for 0 indication on ohms scale. Repeat for each range.
13