TB 9-6625-2357-50
Table 3. Calibration Data Chart for Dial Indication Calibrator (DIC)
Electronic Linear Transducer
ROTATION
TI
Display Unit indication
DIRECTION
Setting
Minimum
Maximum
CW
0.05
0.0499
0.0501
CW
0.1
0.0999
0.1001
CW
0.2
0.1999
0.2001
CW
0.3
0.2999
0.3001
CW
0.4
0.3999
0.4001
CW
0.5
0.4999
0.5001
CW
0.6
0.5999
0.6001
CW
0.7
0.6999
0.7001
CW
0.8
0.7999
0.8001
CW
0.9
0.8999
0.9001
CW
0.95
0.9499
0.9501
CCW
0.5
0.4999
0.5001
CCW
0.0
-0.0001
0.0001
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.