TB 9-6625-2363-35
(5) Repeat technique of (3) (c) above, using the amplitude level and indications
listed in table 4. Measuring receiver will indicate within limits specified in table 4.
Measuring receiver
indications
Test instrument
(dB)
Amplitude
level
Min
Max
7
dBm
6.5
7.5
5
dBm
4.5
5.5
dBm1
-0.5
0.5
0
-10
dBm
-10.5
-9.5
-20
dBm
-20.5
-19.5
-30
dBm
-30.5
-29.5
-40
dBm
-40.5
-39.5
-50
dBm
-50.5
-49.5
-60
dBm
-60.5
-59.5
-70
dBm
-70.5
-69.5
-80
dBm
-80.5
-79.5
-90
dBm
-90.5
-89.5
-100
dBm
-100.5
-99.5
-110
dBm
-110.5
-109.5
1Setup
measuring receiver for tuned RF level cal techniques
and calibrate. Recal as necessary.
(6) Set TI RF OUTPUT to off and disconnect sensor module (11722A). On
measuring receiver exchange sensor module (11792A) for (11722A) and connect sensor
module (11792A) to TI RF OUTPUT.
(7) Press TI pushbuttons as listed in (a) through (d) below:
(a)
Preset.
(b)
(c)
Amplitude, 7, [dBm].
(d)
RF On/Off to on.
(8) Using measuring receiver and RF power measurement techniques, measuring
receiver will indicate within limits specified in first row of table 5.
(9) Repeat technique of (7) (c) above, using the amplitude level and indications
listed in table 5. Measuring receiver will indicate within limits specified in table 5.
(10) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. No adjustments can be made.
CHANGE 1
8