TB 9-6625-336-35
SECTION III
6. Preliminary Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name and item identification number as listed in tables 2 and 3. For the identification of
equipment referenced by item numbers prefixed with A, see table 2, and for prefix B, see
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
NOTE
Unless otherwise specified, verify the result of each test and,
whenever the test requirement is not met, take corrective
action before continuing with the calibration. Adjustments
required to calibrate the TI are included in this procedure.
Additional maintenance information is provided in TM 11 -
6625-493-15 or manufacturer's manual.
NOTE
When indications specified in paragraphs 8 through 11 are not
within tolerance, perform the power supply check prior to
making adjustments. After adjustments are made, repeat
if all other parameters are within tolerance.
NOTE
Unless otherwise specified, all controls and control settings
refer to the TI.
7. Equipment Setup
NOTE
Remove TI protective cover only if necessary to perform
adjustments.
b. Connect autotransformer to a
115-V ac
source and
adjust for a
115-V ac
output.
4