TB 9-6625-357-35
(4) Repeat (3) above with autotransformer output settings of 105 and 125 V.
b. Adjustments. No adjustments can be made.
17. Low
Pass Dial Accuracy and Unity Gain
a. Performance Check
(1) Connect equipment as shown in figure 4, using termination.
Figure 4. Unity
gain check -
equipment setup.
(2) Set CUTOFF FREQ Hz dial to 35 and selector switch to LOW PASS X10K.
(4) Move true rms voltmeter connection from INPUT to OUTPUT. If true rms
voltmeter does not indicate between 0.94 and 1.06 V, perform b (1) below.
(5) Set CUTOFF FREQ Hz dial to 100 and selector switch to HIGH PASS X1. If
true rms voltmeter does not indicate between 0.94 and 1.06 V, perform b (2) below.
(6) Connect frequency counter to INPUT connector.
(7) Set CUTOFF FREQ Hz dial to 60 and selector switch to LOW PASS X100.
PIN: 054515-001