TB 9-6625-388-35
(3) Measure and record resistance value.
(4) Turn decade resistor switches to 9.00 ohms.
(5) Measure and record resistance value. Subtract value recorded in (3) above from
this value. (This is the value of 9 ohms at decade resistor.)
Figure 1. Ratio resistor check - equipment setup.
(7) Position TI controls as listed in (a) through (c) below:
(a) MULTIPLY BY switch to 1/1000.
(b) BA switch to EXT.
(c) 1000-ohm decade switch to 9 and all others to 0.
(8) Adjust decade resistor for 9 ohms.
(9) Adjust resistance measuring system POWER LIMIT control for 1000
MILLIWATTS.
(10) Adjust resistance measuring system RANGE control to 20 volts.
(11) Set resistance measuring system + OFF switch to +.
8