TB 9-6625-396-50
Test instrument G
10a(7) (F)
switch settings
Min
Max
8888
0.879912
0.897688
7777
0.769923
0.785477
6666
0.659934
0.673266
5555
0.549945
0.561055
4444
0.439956
0.448844
3333
0.329967
0.336633
2222
0.219978
0.224422
1111
0.109989
0.112211
b. Adjustments. No adjustments can be made.
10. Final Procedure
a. Deenergize and disconnect all equipment.
b. In accordance with TM 38-750, annotate and affix DA Label 80 (US Army
Calibrated Instrument). When the TI receives limited or special calibration annotate and
affix DA Label 163 (US Army Limited or Special Calibration). When the TI cannot be
adjusted within tolerance, annotate and affix DA Form 2417 (US Army Calibration System
Rejected Instrument).
9