TB 9-6635-208-50
Table 5. Range Check (0.05 to 1.2 Inch-Ounce, 2-Inch Pulley).
Test Instrument
Required
Cardinal
weight
5% of full scale 2% of full scale
point
(oz).
Min
Max
Min
Max
0.25
1/4
0.19
0.31
0.226
0.274
0.50
1/2
0.44
0.56
0.476
0.524
0.75
3/4
0.69
0.81
0.726
0.774
1.00
1
0.94
1.06
0.976
1.024
Table 6. Range Check (0 to 20 Inch-Ounce, 2-Inch Pulley).
Required
Test Instrument
weight
5% of
2% of
Cardinal
(Oz).
full scale
full scale
point
Min
Max
Min
Max
3
3
2
4
2.6
3.4
6
6
5
7
5.6
6.4
9
9
8
10
8.6
9.4
12
12
11
13
11.6
12.4
15
15
14
16
14.6
15.4
18
18
17
19
17.6
18.4
10. Final Procedure
a. Disconnect all equipment and replace TI with in protective cover.
b. In accordance with TM 38-750, annotate and affix DA Label 80 (U.S. Army
Calibration System). When the TI cannot be adjusted within tolerance, annotate and affix
DA Form 2417 (Unserviceable or Limited Use) tag.
6