TB 9-6695-272-35
b. Adjustments
(1) Press ON-STBY/POWER pushbutton to OFF.
(2) Remove top cover.
(3) Remove three bolts that hold module to chassis (fig. 1).
(4) Remove semi rigid coaxial cable (fig. 1).
CAUTION
When performing (5) below, ensure that all cables remain in
place.
(5) Turn TI on side and remove module from chassis (fig. 1).
(6) Remove assemblies A6 and A7 (fig. 1) and place on extender boards.
(7) Install cable from A6 OUTPUT LEVELING to output attenuator (fig. 1).
(8) Press ON-STBY/POWER pushbutton to ON, then press 10, kHz/+dBm,
AMPLITUDE, 13.01, kHz/+dBm pushbuttons.
(10) Connect multimeter between A6TP6 (fig. 2) and chassis ground. Adjust BIAS
ADJ A6R85 (fig. 2)
for 0 V
0.01 V dc
(R).
(11) Remove ground from A6TP5 and adjust DC NULL ADJ A7R92 (fig. 3) for 0 V
+0.01 V dc (R).
(12) Press ON-STBY/POWER pushbutton to OFF.
7