TB 9-6695-281-35
Test instrument
Resistance measuring system indications
TEMPERATURE selector
minus zero resistance value
switch positions
(Ohms)
(inner scale)
Min
Max
90.38 (OHMS) RES.
74.19
74.29
-50
-30
80.51
80.61
-10
86.99
87.09
0
90.33
90.43
+10
93.75
93.85
+30
100.86
100.96
+50
108.34
108.44
+80
120.31
120.41
+100
128.80
128.90
+120
137.73
137.83
+150
151.86
151.96
+200
177.90
178.00
+250
207.95
208.05
+300
242.65
242.75
50.00 (OHMS) RES.
164.45
164.55
+300
+250
138.95
139.05
+200
116.45
116.55
+150
96.45
96.55
+100
78.95
79.05
+50
63.50
63.60
0
49.95
50.05
-50
37.90
38.00
(5) Set TI LEFT and RIGHT switch to RIGHT.
(6) Connect resistance measurement system to TI thermometer test cable pins E
and C Repeat technique of (1) through (4) above.
b. Adjustments. No adjustments can be made.
13. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. When all parameters are within tolerance, annotate and affix DA Label 80 (U.S.
Army Calibrated Instrument). When the TI receives limited or special calibration,
annotate and affix DA Label 163 (U.S. Army Limited or Special Calibration). When the TI
cannot be adjusted within tolerance, repair the TI in accordance with the maintenance
manual. When repair is delayed f or any reason or the TI cannot be repaired with local
resources, annotate and affix DA Form 2417 (U.S. Army Calibration System Rejected
Instrument) and inform the owner/user accordingly in accordance with TB 750-25.
10