TB 9-6695-281-35
10. Thermocouple Thermometer
a. Performance Check
NOTE
A full-scale (red
line)
indication
is
required
for
all
measurements.
(1) Remove TI TESTER STANDARDIZING ONLY cover.
(3) Set TI RESISTANCE and VOLTAGE selector switch to THERMOCOUPLE
BENCH TEST 8 OHMS.
(5) Adjust TI COARSE and FINE rheostats for full-scale (red line) indication on TI
(6) Set TEMPERATURE selector switch to -50 on IRON-CONSTANTAN (degrees)
C outer scale. If digital voltmeter (A1) does not indicate between -3.53 and -3.61 mV dc,
perform b below.
(7) Repeat technique of (4) through (6) above for switch positions listed in table 6.
Digital voltmeter will indicate within limits specified.
Figure 2. Thermocouple thermometer - equipment setup.